Evaluation of iron overload by single voxel MRS measurement of liver T2
Zhiyue J Wang1, John C Haselgrove, Marie B Martin
1Department of Radiology, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA. zjwang@bcm.tmc.edu
Purpose:
To overcome the difficulty of poor signal-to-noise ratio of magnetic resonance imaging (MRI) in evaluating heavy iron overload by using a single voxel magnetic resonance spectroscopy (MRS) technique.
Materials And Methods:
A single voxel STEAM pulse sequence with a minimum TE of 1.5 msec and a sampling volume of 36.6 cm(3) was developed and applied to 1/T2 measurement of the liver in 14 patients with thalassemia whose liver iron concentration was determined through biopsy.
Results:
The iron level ranged from 0.23 to 37.15 mg Fe/g dry tissue with a median value of 18.06. In all cases, strong MR signals were obtained. 1/T2 was strongly correlated with the liver iron concentration (r = 0.95, P < 0.00005).
Conclusion:
The single voxel MRS measurement of T2 in liver iron overload overcomes the difficulty of lack of detectable signals in conventional MRI when the iron level is high. There is an excellent correlation between the iron level and 1/T2.
More Related Videos
09:54Synthesis, Characterization, and Application of Superparamagnetic Iron Oxide Nanoprobes for Extrapulmonary Tuberculosis Detection
Published on: February 16, 2020
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...