Neonatal laboratory blood sampling: comparison of results from arterial catheters with those from an automated
K J Johnson1, G A Cress, N W Connolly
1University of Iowa Hospitals and Clinics, Department of Pediatrics, W226 GH 200 Hawkins Drive, Iowa City, IA 52242, USA. karen-johnson@uiowa.edu
Purpose:
To compare neonatal laboratory results from capillary blood samples drawn using the Tenderfoot automated capillary sampling device with those drawn through arterial catheters.
Design:
Prospective, paired comparisons of laboratory results from capillary and arterial blood.
Sample:
Twenty-one infants being cared for in an NICU and having indwelling arterial catheters through which a variety of predominantly glucose-containing fluids were being administered.
Main Outcome Variables:
Statistical comparisons of paired capillary and arterial results of pH, PO2, PCO2, lactate, sodium, potassium, ionized calcium, and hematocrit.
Results:
No capillary-arterial differences were observed for pH, PCO2, lactate, or sodium. Although capillary results were slightly, but significantly (p < .01), higher for potassium (+0.4 mEq/liter), ionized calcium (+0.47 mg/dl), and hematocrit (+4 percent), these differences fell within acceptable Clinical Laboratories Improvement Act (CLIA) performance criteria. Markedly lower PO2 (-30.2 mmHg) and glucose (-61 mg/dl) values were observed with capillary sampling. With the exception of results for PaO2 and plasma glucose, capillary blood drawn using the Tenderfoot automated device yields laboratory results comparable to those from blood drawn from arterial catheters as assessed by CLIA performance criteria.
Related Concept Videos
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...