Related Experiment Video
Updated: Jul 31, 2026

Detection of a CDH1 Rare Transcript Variant in Fresh-frozen Gastric Cancer Tissues by Chip-based Digital PCR
Published on: February 5, 2018
Detection of sequence variations in the adenomatous polyposis coli (APC) gene using denaturing high-performance
1The Diagnostic Sequencing Laboratory, Department of Molecular and Human Genetics, Baylor College of Medicine, Houston, TX 77030, USA.
Abstract:
We have evaluated the usefulness of denaturing high performance liquid chromatography (dHPLC) for scanning the adenomatous polyposis coli (APC) gene for point mutations, small deletions, and insertions. Our assay consists of 28 sets of primers to amplify the 15 exons of the APC gene. All PCR reactions were amplified simultaneously using the same reaction conditions in a 96-well format and then analyzed by dHPLC, using empirically determined optimum temperatures for partial fragment denaturation. Previously studied DNA specimens from 47 familial adenomatous polyposis (FAP) patients were analyzed by dHPLC and all mutations were correctly identified and confirmed by sequence analysis. This approach identified a single-base substitution in exon 6 and a 2-bp insertion in exon 15 that initially had not been detected by single-strand conformational polymorphism (SSCP) analysis. A novel mutation in exon 15 of the APC gene, 2065delG (codon 689) that had previously been undetected by the protein truncation test (PTT) was also identified by dHPLC. We present our validation studies of dHPLC technology for APC gene analysis in terms of sensitivity and specificity and compare it to current standard scanning technologies including PTT, SSCP, and conformational sensitive gel electrophoresis (CSGE).