Performance of infusion pumps during hyperbaric conditions
Haim Lavon1, Avi Shupak, Dror Tal
1Israel Naval Medical Institute, Israel Defence Forces Medical Corps, Haifa, Israel. dr_lavon@hotmail.com
Anesthesiology
|April 20, 2002
Summary
Infusion pumps show varied accuracy in hyperbaric conditions. The Infutec 520 and Imed 965 performed acceptably, while others failed, highlighting the need for careful equipment selection in hyperbaric facilities.
Area of Science:
- Hyperbaric Medicine
- Biomedical Engineering
- Medical Device Technology
Background:
- Infusion pumps are frequently utilized within hyperbaric chambers.
- Ensuring the accuracy and reliability of these devices under hyperbaric conditions is critical for patient safety.
- This study evaluates the performance of several common infusion pump models.
Purpose of the Study:
- To assess the functional accuracy of volumetric and syringe infusion pumps under simulated hyperbaric conditions.
- To determine the suitability of specific infusion pump models for use in hyperbaric oxygen therapy and decompression protocols.
Main Methods:
- Tested four infusion pump models (Imed 965, Infutec 520, Easy-pump MZ-257, Graseby 3100) under varying hyperbaric pressures (up to 6 atmospheres absolute [ATA]).
- Evaluated pump accuracy at low (12-18 ml/h) and high (60-100 ml/h) infusion rates across different hyperbaric protocols.
- Examined two units of each pump model under simulated hyperbaric oxygen treatment and US Navy decompression tables.
Main Results:
- The Easy-pump MZ-257 and Graseby 3100 exhibited significant functional failures at hyperbaric pressures, rendering them unsuitable.
- The Imed 965 generally maintained acceptable volume deviations (<10%), but showed significant errors (20-40%) at low rates during compression.
- The Infutec 520 demonstrated consistent accuracy (within 10%) across all tested hyperbaric conditions and infusion rates.
Conclusions:
- Commercial infusion pumps display considerable variability in performance and accuracy when operated under hyperbaric conditions.
- Hyperbaric facility staff must rigorously evaluate infusion devices to ensure reliable performance and prevent inaccuracies during patient treatment.
- The Infutec 520 appears to be a suitable option for hyperbaric use, while others require careful consideration or avoidance.