Force reduction by an external hip protector on the human hip after falls
Stanley L Wiener1, Gunnar B J Andersson, Lloyd M Nyhus
1University of Illinois College of Medicine at Chicago, USA.
Clinical Orthopaedics and Related Research
|April 20, 2002
Summary
Hip protectors significantly reduce impact force during falls, transmitting 5% or less to the skin. This study provides evidence that hip protectors are safe and effective in preventing hip fractures in elderly individuals.
Area of Science:
- Biomechanics
- Gerontology
- Orthopedics
Background:
- Hip fractures are a major cause of morbidity and mortality in the elderly.
- Hip protectors are designed to mitigate impact forces during falls.
- Previous studies have not quantitatively assessed impact force reduction in vivo.
Purpose of the Study:
- To quantitatively measure the impact force transmitted to the hip during lateral falls while wearing hip protectors.
- To evaluate the effectiveness of hip protectors in preventing hip injuries.
- To assess the compliance of elderly individuals with hip protector use.
Main Methods:
- Drop weight tests using piezoelectric sensors to measure impact force.
- Volunteer studies involving lateral falls from standing onto a concrete surface with hip protectors.
- In vitro testing with osteoporotic cadaver hips.
- A 1-month compliance study with 106 elderly participants.
Main Results:
- Hip protector pads transmitted only 9% of impact force in drop weight tests.
- In volunteer falls, 5% or less impact force was transmitted to skin sensors.
- No hip fractures occurred in 85 volunteer falls; only transient skin pain was reported.
- Force transmission was below the fracture threshold for osteoporotic cadaver hips.
- Compliance with protector use ranged from 70% to 95% of waking hours.
Conclusions:
- Hip protectors are effective in significantly reducing impact forces during falls.
- The use of hip protectors demonstrates a high safety profile and potential for preventing hip fractures in the elderly.
- Variability in force reduction among individuals was observed, suggesting potential for personalized fitting or pad design improvements.