Related Experiment Videos
The development of an injection-molding process for a polyanhydride implant containing gentamicin sulfate
Jone-Shin Deng1, Marts Meisters, Luk Li
1Hospital Products Division, Abbott Laboratories 97D/AP4, 100 Abbott Park Rd., Abbot Park, IL 60064, USA. DENGJS@HPD.Abbott.com
Abstract:
A production-scale manufacturing process has been developed for polyanhydride/gentamicin sulfate implants for the treatment of osteomyelitis. Gentamicin sulfate was first dried to an acceptable moisture level by using a tumble vacuum dryer. Dried gentamicin sulfate powder and polyanhydride granules were separately fed into the twin-screw extruder at a pre-determined metering rate using a gravimetric feeding device. The extruded molten mixture was solidified to form strands which were subsequently cut into pellets by using a pelletizer. The pellets were characterized with respect to copolymer molecular weight and drug content uniformity. The pellets were later fed into production-scale injection-molding equipment for implant fabrication. The injection-molding cycle was developed and evaluated in terms of cycle reproducibility. Implants were tested and shown to yield an oriented skin-core structure exhibiting a desirable in-vitro drug release profile.