Improved computed radiography image quality from a BaFl:Eu photostimulable phosphor plate
Yasushi Nakano1, Tomonori Gido, Satoshi Honda
1MG Imaging R&D Center, Konica Corporation, Hino-shi, Tokyo, Japan.
Medical Physics
|May 7, 2002
Summary
New photostimulable phosphor (PSP) plates made of BaFI:Eu offer improved computed radiography (CR) system image quality. These plates demonstrate higher detective quantum efficiency (DQE) and lower noise, making them promising for advanced CR applications.
Area of Science:
- Medical Physics
- Materials Science
- Radiography Imaging
Background:
- Photostimulable phosphor (PSP) plates are crucial components in computed radiography (CR) systems.
- Advancements in PSP plate design aim to enhance image quality and diagnostic performance.
- Barium fluoro-halide (BaF(Br,I):Eu) phosphors are commonly used, but new materials are being explored.
Purpose of the Study:
- To evaluate the image quality of newly designed BaFI:Eu phosphor plates for CR systems.
- To compare the performance of BaFI:Eu plates against existing BaF(Br,I):Eu and BaFBr:Eu plates.
- To determine the impact of BaFI:Eu on key image quality metrics like modulation transfer function (MTF), Wiener spectra (WSN), and detective quantum efficiency (DQE).
Main Methods:
- Comparative analysis of PSP plates including BaFI:Eu, BaF(Br,I):Eu, and BaFBr:Eu.
- Evaluation of image quality parameters: presampling modulation transfer function (MTF), normalized Wiener spectra (WSN), and detective quantum efficiency (DQE).
- Measurement of x-ray attenuation coefficients at 60 KeV for different phosphor materials.
Main Results:
- The BaFI:Eu plate demonstrated higher detective quantum efficiency (DQE) compared to the best BaF(Br,I):Eu plate across tested spatial frequencies (0.5, 1.0, 2.0 cycles/mm).
- The improved DQE of BaFI:Eu is attributed to significantly lower normalized Wiener spectra (WSN), indicating reduced image noise.
- BaFI:Eu exhibits higher x-ray attenuation coefficients (27% higher at 60 KeV) due to its iodine halide content, contributing to better x-ray absorption.
Conclusions:
- BaFI:Eu phosphor plates represent a significant advancement in CR technology.
- These plates offer a promising pathway to lower image noise and enhance overall CR system image quality.
- The material's superior x-ray absorption properties contribute to its improved performance metrics.