Experimental evaluation and early clinical results of a new low-profile bileaflet aortic valve
Thomas Walther1, Sven Lehmann, Volkmar Falk
1Klinik für Herzchirurgie, Herzzentrum, Universität Leipzig, Leipzig, Germany. walt@medizin.uni-leipzig.de
Abstract:
We performed an experimental and clinical evaluation of a new low-profile bileaflet aortic valve (Regent, St. Jude Medical Inc., St. Paul, MN, U.S.A.). Common valve sizes were experimentally tested for leakage volume, pressure drop, and transvalvular hemodynamics using a pulse duplicator. Thirty patients (mean age 60 +/- 7 years, predominant valve stenosis n = 25) received the Regent prosthesis for initial clinical evaluation. In vitro evaluation revealed equivalent leakage volumes, larger performance indices (0.552 versus 0.513), and lower pressure drops in comparison to SJM hemodynamic plus valve controls. Clinically, 21 mm (n = 9), 23 mm (n = 12), and 25 mm (n = 9) valves were implanted with no significant perioperative complications. Echocardiography revealed low transvalvular flow velocities (2.2 +/- 0.4 m/s) and low pressure gradients (20 +/- 6 mm Hg) postoperatively and at 6 months follow-up. In vitro testing and early clinical results are promising; however, long-term performance has to be proven.
More Related Videos
Related Concept Videos
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation III: Medical Management