Microdialysis of Ethanol During Operant Ethanol Self-administration and Ethanol Determination by Gas Chromatography
Published on: September 5, 2012
Optimization of solid-phase microextraction conditions for gas chromatographic determination of ethanol and other
Dariusz Zuba1, Andrzej Parczewski, Manfred Reichenbächer
1Institute of Forensic Research, Westerplatte 9, PL 31-033, Krakow, Poland.
Abstract:
A procedure for the determination of acetaldehyde, acetone, methanol, ethanol, 1-propanol and 2-propanol in blood was developed. Separation of analytes was carried out on DB-wax capillary column (l = 30 m, I.D. = 0.32 mm, dF = 0.5 microm) at 40 degrees C, hydrogen was used as a carrier gas (at 30 kPa) and FID as a detector. Quantification was performed with the use of 2-butanol as an internal standard. Headspace solid-phase microextraction was applied as the sample preparation technique. The usefulness of most commercially available fiber coatings was checked and 65 microm Carbowax/DVB proved most effective. Microextraction was carried out from the headspace at 60 degrees C for 10 min. The sample was stirred at 750 rpm. In order to improve the extraction efficiency of analytes, salting-out agents were also applied. Potassium carbonate turned out to be the most efficient. A 1.0-g amount of this salt and 0.1 ml of I.S. were added to 0.5 ml of sample. Validation of the worked-out method was performed. For each analyte, the limits of detection and quantification, linearity, working range, accuracy and precision were determined or tested.
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...