Related Experiment Video
Updated: Oct 1, 2026

Dynamic Multiparameter Platelet Function Assessment Using a Capacitive Biosensor
Published on: May 2, 2025
Rational use of the PFA-100 device for screening of platelet function disorders and von Willebrand disease
Y Buyukasik1, S Karakus, H Goker
1Department of Hematology, Hacettepe University Medical School, Ankara, Turkey. ybuyukas@hacettepe.edu.tr
Abstract:
Two hundred and five patients referred for evaluation of platelet functions and 126 healthy controls were tested with the PFA-100 instrument. A cut-off value of 150 s for collagen/epinephrine (CEPI) closure time (CT) produced most acceptable sensitivity (90%), specificity (85.2%), and positive (82.6%) and negative (91.6%) predictivity values for screening of platelet function disorders and von Willebrand disease (vWD). All patients with vWD and Glanzmann thrombasthenia could be detected by PFA-100. Both CEPI and collagen/adenosine diphosphate (CADP) CTs were elevated in all of these cases. Sensitivity of the device was 81.6% for patients with platelet secretion defects. CADP CT was normal in 63.9% of the patients in this subgroup. Specificity (47%) and positive predictivity (57%) of the instrument were diminished in patients with low hemoglobin concentrations. Depending on the results, an algorithm was developed for screening of platelet function disorders and vWD with PFA-100.

