A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
Decision limit for troponin I and assay performance
Paul Sheehan1, John Blennerhassett, Samuel D Vasikaran
1Division of Laboratory Medicine, Royal Perth Hospital, Australia. paulsheehan@health.wa.gov.au
Assay imprecision with the Bayer ACS:180 troponin I assay can lead to inconsistent risk assessment for acute myocardial infarction (AMI) patients. Careful attention to sample quality is vital for accurate troponin I measurement.
Area of Science:
- Clinical Chemistry
- Cardiovascular Diagnostics
- Biomarker Assay Performance
Background:
- Redefinition of acute myocardial infarction (AMI) emphasizes troponin diagnostics.
- New guidelines lower diagnostic cut-offs to the 99th percentile, requiring high assay precision (10% CV).
- The Bayer ACS:180 troponin I assay is evaluated for risk stratification at a 0.1 microg/L cut-off.
Purpose of the Study:
- Assess the appropriateness of the Bayer ACS:180 assay at a 0.1 microg/L cut-off for individual patient risk stratification.
- Determine the functional sensitivity of the troponin I assay.
- Evaluate the impact of sample quality on assay performance.
Main Methods:
- Analysis of assay imprecision at the 0.1 microg/L decision limit.
- Determination of functional sensitivity at 10% inter-assay coefficient of variation (CV).
- Assessment of sample quality by monitoring duplicate precision checks.
Main Results:
- Assay imprecision at 0.1 microg/L could alter risk group assignment for 8% of patients upon re-analysis.
- Functional sensitivity of the ACS:180 troponin I assay was found to be 0.27 microg/L (at 10% CV).
- Poor sample quality, typically fibrin particles, caused 0.39% of routine samples to fail precision checks.
Conclusions:
- Inconsistent risk assessment can occur in a small but significant proportion of cases when using the Bayer assay at 0.1 microg/L.
- Maintaining high sample quality is crucial for reliable troponin I measurements.
- The current performance characteristics of the assay may limit its utility for precise risk stratification at the specified cut-off.