Related Experiment Video
Updated: Oct 1, 2026

Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
Quality control materials for the determination of trace elements in airborne particulate matter
J Sysalová1, J Kucera, B Kotlík
1Institute of Analytical Chemistry, Academy of Sciences of the Czech Republic, Vídenská 1083, 142 20 Prague 4, Czech Republic. sysalova@cgu.cz
Abstract:
The development of quality control materials for the determination of selected trace elements in air pollution studies is described. Three types of test samples were prepared for proficiency testing: (1) filters loaded with PM10 fraction of urban air particulate matter (APM) using high-volume air samplers, which were subsequently divided into smaller sections, (2) a bulk sample of APM collected in an automobile tunnel in Prague, and (3) simulated air filters loaded with APM using a wet deposition process. Homogeneity of the test samples was studied using instrumental neutron activation analysis, proton induced X-ray emission and atomic absorption spectrometry, and inductively coupled plasma optical emission spectrometry and mass spectrometry. Sufficiently homogeneous samples were prepared by all three procedures. The simulated air filters appeared to be the most suitable test samples for proficiency testing.
Related Concept Videos
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Atomic Absorption Spectroscopy: Lab