Determination of accuracy in neonates for non-invasive blood pressure device using an improved algorithm
Related Experiment Video
Updated: Oct 1, 2026

Hemodynamic Precision in the Neonatal Intensive Care Unit using Targeted Neonatal Echocardiography
Published on: January 27, 2023
Robert M Nelson1, Andrea D Stebor, Catherine M Groh
1Division of Neonatology, Department of Pediatrics, University of South Florida, Tampa 33606-3309, USA. rnelson@med.usf.edu
Background:
New software (SuperSTAT(R) algorithm) with enhancements aimed at shorter determination times was developed for a non-invasive blood pressure (NIBP) device and a clinical evaluation was conducted to verify accuracy.
Objective:
To determine the accuracy of the new algorithm according to ANSI/AAMI SP10-1992 and SP10A-1996 American National Standard for Electronic or Automated Sphygmomanometers.
Methods:
The blood pressure values obtained from the test device were compared to the intra-arterial blood pressure reference standard (IBP).
Results:
The NIBP and IBP comparisons for systolic, diastolic, and mean arterial pressure met the 1992 ANSI/AAMI accuracy standards by being within a mean difference of +/- 5 mmHg and standard deviation of < or = 8 mmHg.
Conclusion:
Non-invasive blood pressure determinations taken with the new algorithm, developed to provide greater patient comfort due to faster determinations, were accurate when compared to neonatal IBP.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure