Related Experiment Videos
Direct amperometric determination of lactate at a carbon fiber bundle microdisk electrode by capillary zone
Qian Dong1, Rui Dong, Mingliang Jin
1School of Chemistry and Chemical Engineering, and State Key Laboratory of Microbial Technology, Shandong University, Jinan 250100, PR China.
Abstract:
Capillary zone electrophoresis was employed for the determination of lactate using end-column amperometric detection at a carbon fiber bundle microdisk electrode. The optimum conditions of separation and detection are 3.6 x 10(-3) mol/l Na(2)HPO(4)-1.4 x 10(-3) mol/l NaH(2)PO (pH 7.2) for the buffer solution, 18 kV for the separation voltage and 1.60 V versus the saturated calomel electrode for the detection potential. The limit of detection is 7.6 x 10(-7) mol/l or 1.7 fmol (S/N=3) and the linear range is 1.7 x 10(-6)-8.2 x 10(-4) mol/l for the injection voltage of 6 kV and injection time of 5 s. The RSD is 1.8% for the migration time and 3.3% for the electrophoretic peak current. The method was applied to the determination of lactate in human saliva. The recovery of the method is between 95 and 109%.