Validation up-date.
1Blood Pressure Unit and ADAPT Centre, Beaumont Hospital, Dublin 9, Ireland. eobrien@iol.ie
Blood Pressure Monitoring
|June 11, 2002
Summary
New validation protocols simplify blood pressure device evaluation. The European Society of Hypertension (ESH) International Protocol streamlines testing for automated and semi-automated blood pressure monitors.
Area of Science:
- Cardiovascular technology
- Medical device evaluation
- Hypertension research
Background:
- Automated and semi-automated blood pressure devices are increasingly marketed.
- Existing validation protocols (AAMI, BHS) are difficult to fulfill.
- A need exists for standardized, practical evaluation criteria.
Purpose of the Study:
- To introduce the European Society of Hypertension (ESH) International Protocol for blood pressure device validation.
- To simplify and rationalize existing validation procedures.
- To ensure reliable evaluation of automated and semi-automated blood pressure monitors.
Main Methods:
- Based on data from 19 validation studies using AAMI and BHS protocols.
- Critical assessment and rationalization of previous validation methodologies.
- Elimination of pre-validation phases and simplification of observer recruitment/training.
Main Results:
- Simplified validation procedures without compromising rigor.
- Reduced number of subjects and relaxed blood pressure range requirements.
- Early elimination of non-viable devices, improving efficiency.
Conclusions:
- The ESH International Protocol offers a more practical approach to blood pressure device validation.
- The new protocol maintains the merits of earlier, more complex methods.
- Facilitates reliable assessment for purchasers of blood pressure monitoring devices.