Cleveland clinic CorAide blood pump circulatory support without anticoagulation

Yoshie Ochiai1, Leonard A R Golding, Alex L Massiello

  • 1Department of Biomedical Engineering, Lerner Research Institute, The Cleveland Clinic Foundation, Ohio 44195, USA.

ASAIO Journal (American Society for Artificial Internal Organs : 1992)
|June 13, 2002
PubMed

Insights

The CorAide left ventricular assist system demonstrated safe operation in calves with minimal or no anticoagulation. This centrifugal pump showed stable hemodynamics and clean blood pump surfaces, indicating reduced thrombotic risk.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Devices
  • Medical Technology

Background:

  • Left ventricular assist devices (LVADs) are crucial for managing heart failure.
  • Anticoagulation therapy is often required to prevent thrombosis with LVADs, posing bleeding risks.
  • The CorAide system utilizes a noncontact, suspended rotating assembly, potentially reducing thrombogenicity.

Purpose of the Study:

  • To evaluate the biologic effects and safety of the CorAide LVAD in a chronic animal model.
  • To assess the necessity and efficacy of anticoagulation therapy with the CorAide system.
  • To determine the thrombotic potential of the CorAide pump without continuous anticoagulation.

Main Methods:

  • Chronic implantation of the CorAide centrifugal pump in six calves.
  • Varying durations of heparin administration post-surgery, with extended periods without anticoagulation.
  • Hemodynamic monitoring (pump flow, mean arterial pressure) and post-mortem examination for thrombus, device integrity, and adverse events.

Main Results:

  • Stable hemodynamics were maintained across all animals (mean pump flow 5.6 L/min, mean arterial pressure 100 mm Hg).
  • Blood pump surfaces remained free of thrombus in all six calves.
  • No mechanical failures, bleeding complications, or device infections occurred; one case of renal infarction was noted.

Conclusions:

  • The CorAide left ventricular assist system can be safely operated with minimal or no anticoagulation therapy.
  • The noncontact design of the CorAide pump appears to mitigate thrombus formation.
  • These findings support the potential for reduced anticoagulation burden in patients using the CorAide LVAD.