[Digital full field mammography: comparison between radiographic direct magnification and digital monitor zooming]
U Fischer1, F Baum, S Obenauer
1Abteilung Diagnostische Radiologie, Georg-August-Universität Göttingen, Robert-Koch-Strasse 40, 37075 Göttingen. uwe.fischer-weende@t-online.de
Purpose:
Our goal was to compare digital magnification mammograms with images zoomed from the digital contact mammogram in patients with microcalcifications.
Patients And Methods:
Fifty-five patients with 57 microcalcification clusters were evaluated with a FFDM system (Senographe 2000D, GE). In addition to a digital contact mammogram, a digital direct magnification mammogram (factor 1.8 [MAG1.8]) and an image zoomed from the contact mammogram with a magnification factor of 1.8 [ZOOM1.8] were obtained in each patient. The image quality (perfect = 5 points to inadequate = 1 point) and the characterization of microcalcifications (BI-RADS 2-5) were evaluated by 4 readers. The results were compared to histopathologic findings in 35 patients (37 lesions) and follow-up in 20 patients.
Results:
Histopathology revealed 16 benign and 21 malignant lesions. 20 patients had benign changes verified by long-term follow-up. Image quality of direct magnification FFDM was assessed superior (4.44 points) to zoomed images (4.14 points). Sensitivity was superior for direct magnification (97.5%) in comparison to the zoomed images (96.3%). However, specificity (MAG1.8: 34.3%, ZOOM1.8: 40%), PPV (MAG1.8: 47.5%, ZOOM1.8: 49.8%) and accuracy (MAG1.8: 58.1%, ZOOM1.8: 61.2%) were better with zooming technique. Deviation steps from best BI-RADS assessment were 0.45 for MAG1.8 and 0.44 for ZOOM1.8.
Conclusions:
In patients with mammographic microcalcifications, monitor zooming of the digital contact mammogram is equivalent to direct magnification FFDM. Therefore, monitor zooming allows a reduction of the radiation exposure and an optimization of the work-flow.
Related Concept Videos
Three-Dimensional Microscopy in Microbiology
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...