Protocol for Production of a Genetic Cross of the Rodent Malaria Parasites
Published on: January 3, 2011
Artefactually-normal automated platelet counts due to malaria-infected RBC
G Crabbe1, M Van Poucke, B Cantinieaux
1CHU Saint-Pierre, ULB. VUB, 322 rue Haute Brussels 1000, Belgium.
Clinical and Laboratory Haematology
|June 18, 2002
Summary
Automated blood analyzers can miscount platelets due to red blood cell fragments, especially in malaria cases. Microscopic examination and immunological platelet counts are crucial for accurate diagnosis when automated systems show discrepancies.
Area of Science:
- Hematology
- Clinical Pathology
- Parasitology
Background:
- Automated blood analyzers utilize impedance and optical technologies for platelet counting.
- Protein aggregates, cellular fragments, and other particles can interfere with these automated counts, leading to inaccurate platelet values.
- Pseudo-thrombocytopenia or falsely elevated platelet counts can occur due to such interferences.
Observation:
- A case of treated malaria presented with a normal platelet count on automated analysis.
- Examination of the blood smear revealed small red blood cells infected with Plasmodium falciparum trophozoites.
- These infected red blood cells interfered with the platelet count in the Cell Dyn 4000 analyzer.
Findings:
- The Cell Dyn 4000 exhibited distinct interference patterns in both its impedance and optical platelet counts due to infected red blood cells.
- The automated analysis initially suggested thrombocytopenia (low platelet count).
- Confirmation of thrombocytopenia was achieved through a phase-contrast microscopic platelet count, a more reliable method in this context.
Implications:
- This case highlights the potential for parasitic infections, specifically malaria, to cause significant interference in automated platelet counting.
- Systematic examination of stained blood films and employing reference methods like immunological platelet counts or phase-contrast microscopy are essential for accurate diagnosis in suspected interference cases.
- Accurate platelet counts are critical for managing conditions like malaria and preventing complications associated with thrombocytopenia.