Related Experiment Video
Updated: Sep 3, 2026

High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
Continuous on-line measurement of absolute left ventricular volume by transcardiac conductance: angiographic
Eva M Staal1, Paul Steendijk, Gerhard Koning
1Department of Cardiology, Leiden University Medical Center, The Netherlands.
Objective:
Validation of the transcardiac conductance method for continuous, on-line measurement of absolute left ventricular volume by comparison with biplane angiography.
Design:
Controlled, prospective animal study.
Setting:
Catheterization laboratory of the Leiden University Medical Center.
Subjects:
Six anesthetized sheep.
Interventions:
Subjects were studied at baseline, during infusion of dobutamine, and during volume loading and beta blockade. In a pilot experiment, a coronary artery was occluded by a balloon, and the behavior of the transcardiac conductance signals during ischemia was tested.
Measurements And Main Results:
Calibration factors alpha and V(p) were determined by thermodilution and hypertonic saline dilution, respectively. Calibrated transcardiac conductance volume was compared with angiographic volume in four different hemodynamic conditions, and transcardiac conductance measurements were registered during a period of ischemia. Results showed a good linear correlation between transcardiac conductance and angiographic volume (r =.77, p <.01) with an intercept of 12.5 +/- 5.6 mL (interanimal variability, 17.8 mL) and a slope of 1.49 +/- 0.15 (interanimal variability, 0.34). Mean alpha and V(p) were 0.12 +/- 0.01 (interanimal variability, 0.07) and 104 +/- 3 mL (interanimal variability, 38 mL), respectively. V(p) did not vary significantly between conditions, and alpha varied only during propranolol (p =.04). Transcardiac conductance enabled immediate visualization of acute left ventricular volume changes during coronary occlusion in a pilot experiment.
Conclusions:
Transcardiac conductance is a method to register an on-line, continuous, left ventricular volume signal, which correlates well with angiography. However, calibration factors need to be determined in individual subjects. The method appears promising to monitor absolute volume in the intensive care unit.