Semiautomatic and robotic methods for determining serum haptoglobin levels.

Joseph E. Smith1, Patricia S. Chavey, Gordon A. Andrews

  • 1Department of Diagnostic Medicine/Pathobiology, College of Veterinary Medicine, Kansas State University, Manhattan, KS 66506, USA.

Veterinary Clinical Pathology
|June 21, 2002
PubMed
Summary

This study adapted the Makimura and Suzuki serum haptoglobin assay for high-throughput analysis. New semiautomated and robotic methods offer efficient and reliable haptoglobin determination in clinical settings.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A method for freezing canine erythrocytes for blood typing purposes.

Veterinary clinical pathology·1994
Same author

An improved ferritin assay for canine sera.

Veterinary clinical pathology·1992
See all related articles

Area of Science:

  • Clinical Chemistry
  • Biochemical Assays
  • Laboratory Automation

Background:

  • The Makimura and Suzuki assay is a standard method for serum haptoglobin determination.
  • High-throughput analysis is crucial for managing large numbers of clinical samples.

Purpose of the Study:

  • To adapt the Makimura and Suzuki serum haptoglobin assay for high-throughput analysis.
  • To evaluate the performance of semiautomated and robotic adaptations of the assay.

Main Methods:

  • Adapted the Makimura and Suzuki serum haptoglobin assay into a one-tube format.
  • Implemented semiautomated and robotic platforms for assay execution.
  • Correlated results with the original two-tube method.

Main Results:

Related Experiment Videos

  • The adapted assay allows for the determination of haptoglobin in several hundred serum samples daily.
  • Both semiautomated and robotic methods showed strong correlation with the original assay.
  • Within-assay variability was 4.4% (semiautomated) and 4.8% (robotic).
  • Day-to-day variability averaged 6.0% (semiautomated) and 5.5% (robotic).
  • Both adapted assays yielded higher y-intercepts, necessitating updated reference values.

Conclusions:

  • The one-tube adaptation of the Makimura and Suzuki serum haptoglobin assay is suitable for high-throughput analysis.
  • Semiautomated and robotic methods provide reliable and efficient haptoglobin determination.
  • Updated reference ranges are required for the adapted assay methods.