X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Impact of JPEG lossy image compression on quantitative digital subtraction radiography
1University of Ljubljana, Faculty of Medicine, Department of Stomatology, Slovenia.
Objectives:
The aim of the study was to evaluate the impact of JPEG lossy image compression on the estimation of alveolar bone gain by quantitative digital subtraction radiography (DSR).
Methods:
Nine dry domestic pig mandible posterior segments were radiographed three times ('Baseline', 'No change', and 'Gain') with standardized projection geometry. Bone gain was simulated by adding artificial bone chips (1, 4, and 15 mg). Images were either compressed before or after registration. No change areas in compressed and subtracted 'No change-Baseline' images and bone gain volumes in compressed and subtracted 'Gain-Baseline' images were calculated and compared to the corresponding measurements performed on original subtracted images.
Results:
Measurements of no change areas ('No change-Baseline') were only slightly affected by compressions down to JPEG 50 (J50) applied either before or after registration. Simulated gain of alveolar bone ('Gain-Baseline') was underestimated when compression before registration was performed. The underestimation was bigger when small bone chips of 1 mg were measured and when higher compression rates were used. Bone chips of 4 and 15 mg were only slightly underestimated when using J90, J70, and J50 compressions before registration.
Conclusions:
Lossy JPEG compression does not affect the measurements of no change areas by DSR. Images undergoing subtraction should be registered before compression and if so, J90 compression with a compression ratio of 1:7 can be used to detect and measure 4 mg and larger bone gain.
More Related Videos
Related Concept Videos
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...