Determination of WR-1065 and WR-33278 by liquid chromatography with electrochemical detection
James A Pendergrass1, Venkataraman Srinivasan, K Sree Kumar
1Armed Forces Radiobiology Research Institute, Radiation Medicine Department, Bethesda, MD 20889-5603, USA. pendergrass@afrri.usuhs.mil
Abstract:
A liquid chromatographic method using electrochemical detection is presented for measuring the thiol WR-1065 12-(3-aminopropylamino) ethanethiol] and its symmetrical disulfide WR-33278 [NH2(CH2)3NHCH2CH2S]2. WR- 1065 is the active, radioprotective drug derived from the phosphorothioate pro-drug WR-2721 (amifostine). External standard curves for both compounds were linear over the range of 40-200 pmol injected (r2 = 0.999 and 0.996 for the thiol and disulfide, respectively). The detection and quantitation limits for WR-1065 were 9 and 18 pmol, respectively, whereas the corresponding values for WR-33278 were 30 and 59 pmol, respectively. Within- and between-day determinations of measurement Vision and accuracy for both compounds validated the suitability of this assay method.
More Related Videos
08:54Determining Four Components in a Lipid Nanoparticle RNA Delivery System by Liquid Chromatography Combined with Evaporative Light Scattering Detector
Published on: May 30, 2025
10:34Characterization and Application of Passive Samplers for Monitoring of Pesticides in Water
Published on: August 3, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
High-Performance Liquid Chromatography: Instrumentation
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process