Related Experiment Video
Updated: Sep 30, 2026

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Modeling the effect of progressive endotracheal tube occlusion on tidal volume in pressure-control mode
1Department of Anesthesia and Critical Care, University of Chicago, Chicago, IL 60637, USA. atung@airway.uchicago.edu
Unlabelled:
A recognized hazard of prolonged endotracheal intubation is progressive airway occlusion resulting from deposition of secretions on the inner surface of the endotracheal tube (ETT). When volume-controlled ventilation is used, progressive ETT occlusion may be detected by monitoring the difference between peak and plateau airway pressures. In pressure-controlled modes, however, inspiratory airway pressures are preset and thus cannot act as a warning indicator. Instead, changes in delivered tidal volumes may aid the diagnosis of ETT occlusion. To determine whether tidal volume monitoring effectively detects progressive ETT occlusion, we mathematically modeled the response of a ventilator operating in pressure-controlled mode to increasing airway resistance. To corroborate our model, we then bench-tested the Siemens 300 and Puritan-Bennett 7200 ventilators by using a test lung and a series of ETTs ranging in size from 9.0 to 3.5 mm inner diameter to simulate progressive occlusion. We found that when pressure-controlled mode was used, progressive ETT occlusion did not reduce delivered tidal volumes until occlusion was nearly complete. We conclude that prolonged use of pressure-controlled mode may allow significant ETT obstruction to build up undetected, risking complete ETT occlusion and complicating the perioperative care of patients ventilated with this mode.
Implications:
Although increasing airway pressures during volume-controlled ventilation allow early recognition of endotracheal tube (ETT) obstruction, airway pressures with pressure-controlled ventilation are fixed. We found during tests of two intensive care unit ventilators that although ETT obstruction reduces delivered tidal volumes during pressure-controlled ventilation, reductions do not occur until occlusion is advanced.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation I: Indication and Settings
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...