Are circular external fixators weakened by the use of hemispheric washers?
Denis J Marcellin-Little1, Simon C Roe, Gian Luca Rovesti
1Department of Clinical Sciences, College of Veterinary Medicine, North Carolina State University, Raleigh, NC 27606, USA.
Objectives:
To compare the axial mechanical stability of 3 circular external fixators systems with and without hemispheric washers.
Study Design:
Experimental study.
Methods:
The axial stiffness and load necessary to produce 0.5 and 1 mm of displacement of 10 circular external fixator constructs from 3 manufacturers were tested on a materials testing machine. The constructs tested included the Small Bone fixator (SBF; Hofmann S.a.S., Monza, Italy), the IMEX ring fixator (IMEX Inc., Longview, TX), and the Multiplanar C-Fix (MCF; PanVet Distribuzione, Seriate, Italy). Five configurations were tested for each construct: (1) conventional nut fixation, (2) hemispheric washer fixation with connecting rods offset by 0, (3) 1, and (4) 2 holes, and (5) with a ring placed at maximum angulation.
Results:
The loads resisted at 0.5 and 1 mm of displacement did not differ when frame configurations were compared (P =.25733 and.33769, respectively). The linear stiffness of the following configurations were decreasingly stiff: standard constructs, hemispheric washers with connecting rods perpendicular to rings, hemispheric washers with connecting rods offset by 1 hole, hemispheric washers with connecting rods offset by 2 holes, and ring offset in relation to bone model. The SBF constructs tested were 34% and 41% more rigid than the IMEX and MCF constructs tested despite the larger diameter of the connecting rods for the IMEX frames (6 mm) compared with the SBF frames (5 mm). The IMEX constructs tested were 6% more rigid than the MCF constructs tested.
Conclusions:
Adding hemispheric washers and angling connecting rods in relation to rings did not influence the loads resisted at 0.5 and 1 mm displacement but decreased construct stiffness.
Clinical Relevance:
The use of hemispheric washers had minor effects on the biomechanical performance of fixator frames tested in this study when used to angle a ring in relation to connecting rods for circular external fixators.
More Related Videos
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
Related Concept Videos
Deformation in a Circular Shaft
Circular Shaft - Stresses in Linear Range
Stress Concentrations in Circular Shafts
Plastic Deformation in Circular Shafts
Residual Stresses in Circular Shafts
Torsion of Noncircular Members