Comparisons of point and average organ dose within an anthropomorphic physical phantom and a computational model of
J B Sessions1, J N Roshau, M A Tressler
1Department of Nuclear and Radiological Engineering, University of Florida, Gainesville 32611-8300, USA.
Insights
This study developed a newborn phantom and simulation model to measure organ radiation dose during pediatric X-rays. MOSFET measurements closely matched Monte Carlo simulations, aiding accurate internal organ dose assessment for medical risk evaluation.
Area of Science:
- Medical Physics
- Radiological Dosimetry
- Pediatric Imaging
Background:
- Pediatric radiographic examinations provide diagnostic benefits but carry radiation risks.
- Accurate internal organ dose assessment is crucial for pediatric medical risk evaluation.
- Existing methods require improved tools for measuring radiation dose in children.
Purpose of the Study:
- To develop and validate a physical phantom and Monte Carlo simulation model for newborn organ dose estimation.
- To compare MOSFET dosimeter measurements with Monte Carlo simulations for pediatric radiographic exams.
- To establish scaling factors for adjusting point dose measurements to volume-averaged organ dose.
Main Methods:
- Developed a physical newborn phantom using tissue-equivalent materials.
- Utilized twenty MOSFET dosimeters placed at organ centroid locations.
- Employed MCNP Monte Carlo simulations for dose calculations.
- Validated measurements against simulations for head, chest, and abdomen exams.
Main Results:
- MOSFET-derived organ doses showed reasonable agreement with Monte Carlo estimates.
- Point-to-organ dose scaling factors (SF(POD)) were determined for clinical use.
- In-field point dose measurements require adjustments within 10% for volume-averaged dose.
- Out-of-field organs and walled organs (stomach, bladder) showed larger dose discrepancies.
Conclusions:
- The developed phantom and simulation model provide a reliable method for pediatric organ dose assessment.
- MOSFET measurements, when adjusted using SF(POD), accurately estimate volume-averaged organ dose.
- This tool aids in balancing diagnostic benefits against radiation risks in pediatric radiography.
Abstract:
Pediatric radiographic examinations yield medical benefits and/or diagnostic information that must be balanced against potential risk from patient radiation exposure. Consequently, clinical tools for measuring internal organ dose are needed for medical risk assessment. In this study, a physical phantom and Monte Carlo simulation model of the newborn patient were developed based upon their stylized mathematical expressions (ORNL and MIRD model series). The physical phantom was constructed using tissue equivalent substitutes for soft tissue, lung, and skeleton. Twenty metal-oxide-semiconductor field effect transistor (MOSFET) dosimeters were then inserted at three-dimensional positions representing the centroids of organs assigned in the ICRP's definition of the effective dose. MOSFET-derived point estimates of organ dose were shown to be in reasonable agreement with Monte Carlo estimates for representative newborn head, chest, and abdomen radiographic exams. Ratios of average organ dose assessed via MCNP simulations to the MOSFET-derived point doses (point-to-organ dose scaling factors, SF(POD)) are tabulated for subsequent use in clinical irradiations of the newborn phantom/MOSFET system. Values of SF(POD) indicate that MOSFET measurements of point dose for in-field exposures need to be adjusted only to within 10% to report volume-averaged organ dose. Larger adjustments to point doses are noted for organs out-of-field. For walled organs, point estimates of organ dose at the content centroid are shown to underestimate the average wall dose when the organ is within the primary field: SF(POD) of 1.19 for the stomach (AP chest exam), and SF(POD) of 1.15 for the urinary bladder (AP abdomen exam).
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Drug Dosing: Infants and Children