SoftCopy Display Quality Assurance Program at Texas Children's Hospital
1Edward B. Singleton Diagnostic Imaging Services, Texas Children's Hospital, Houston 77030-2399, USA. ckly@TexasChildrensHospital.org
A quality assurance program for Picture Archiving and Communication Systems (PACS) displays is crucial for medical imaging. Texas Children's Hospital is developing a program to assess monitor performance, calibration frequency, and useful life for optimal image quality.
Area of Science:
- Medical Imaging Technology
- Radiology Quality Assurance
Background:
- Growing reliance on Picture Archiving and Communication Systems (PACS) necessitates robust workstation display quality assurance.
- No universally accepted PACS quality assurance program exists, though efforts like DICOM Working Group 11 and AAPM Task Group 18 are underway.
Purpose of the Study:
- To develop and evaluate a quality assurance program for PACS workstation displays at Texas Children's Hospital (TCH).
- To collect data for determining optimal calibration frequency, monitor lifespan, and performance limits.
Main Methods:
- Monthly luminance measurements using Society of Motion Picture and Television Engineers (SMPTE) test patterns.
- Subjective evaluations of display sharpness, distortion, and artifacts.
- Vendor calibration every three months, including checks of display parameters and Grayscale Display Function (GSDF) Index.
Main Results:
- Data collection on individual monitor luminance functions is ongoing.
- Service calls are initiated when luminance deviates by more than 10% from arbitrary limits.
- Ongoing analysis of luminance drift, calibration frequency, and monitor lifespan.
Conclusions:
- Establishing a comprehensive QA program is vital for maintaining diagnostic image quality from soft-copy displays.
- Data-driven insights are essential for defining appropriate testing intervals and monitor performance standards.
- Continuous monitoring and calibration are key to ensuring the reliability and longevity of PACS displays.