Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Quantitative evaluation of acquisition parameters in three-dimensional imaging with multidetector computed tomography
Dong-Ook Kim1, Hee-Joung Kim, Haijo Jung
1BK21 Project for Medical Sciences, Research Institute of Radiological Science, Seoul, Korea.
Abstract:
This study evaluated the quantitative measurements of three-dimensional (3D) volume images using multidetector row computed tomography (MDCT) and one skull specimen. Twenty-one linear distances were measured five times each by vernier caliper. A dry human skull was imaged with MDCT for various acquisition parameters at slice thicknesses of 1.25, 2.50, 3.75, and 5.00 mm for the different acquisition modes of axial and spiral scan. The distance of each corresponding item displayed on 3D rendered images was measured seven times by an uninvolved observer using a 3D software tool. Data analysis was performed to determine if there were any statistically significant differences in acquisition parameters. No significant image differences were found among the scan modes for each slice thickness. For a given scan mode, acquisition slice thickness was the important factor for quantitative measurement of 3D rendered CT images.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography