Effect of haptic angulation on posterior capsule opacification in modern foldable lenses with a square, truncated
Josef M Schmidbauer1, Marcela Escobar-Gomez, David J Apple
1Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina 29425-5536, USA.
Purpose:
To analyze the effect of different haptic angulations on posterior capsule opacification (PCO) in a modern foldable intraocular lens (IOL) with a square-edged optic designed to reduce the incidence of PCO.
Setting:
Center for Research on Ocular Therapeutics and Biodevices, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA.
Methods:
Ten Dutch Belted, serum Pasteurella-free pigmented rabbits of the same age and sex had bilateral phacoemulsification with endocapsular IOL implantation. The eyes were implanted with Centerflex IOLs (Rayner) with haptic angulations of 0 degree (n = 8), 5 degrees (n = 4), 10 degrees (n = 4), or 15 degrees (n = 4).
Results:
There was no statistical difference in central PCO, peripheral PCO, and measured IOL decentration among the angulation groups.
Conclusion:
With the Centerflex IOL, haptic angulation did not seem to be a significant factor influencing PCO. It appears that the barrier effect of the IOL's truncated, square-edged optic overrides the angulation factor in preventing PCO.