Related Experiment Video
Updated: Sep 8, 2026

Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
The effects of acute and chronic amiodarone on activation patterns and defibrillation threshold during ventricular
Jian Huang1, Jonathan L Skinner, Jack M Rogers
1Cardiac Rhythm Management Laboratory, Department of Medicine, University of Alabama at Birmingham, 1670 University Boulevard, Birmingham, AL 35294-0019, USA. jh@crml.uab.edu
Objectives:
The goal of this study was to evaluate the effects of acute and chronic amiodarone on activation patterns during ventricular fibrillation (VF), ventricular effective refractory period (VERP) and defibrillation threshold (DFT).
Background:
Acute and chronic amiodarone may act through different mechanisms.
Methods:
The VERP, VF activation patterns and DFT were determined in 24 dogs. Twelve dogs received acute intravenous amiodarone (10 mg/kg, n = 6) or saline (n = 6), and 12 dogs received chronic oral amiodarone (20 mg/kg/day, n = 6) or placebo (n = 6). Epicardial VF activation patterns were recorded with 504 electrodes. Quantitative descriptors of VF were calculated.
Results:
The DFT was unchanged by acute or chronic amiodarone. Although chronic amiodarone significantly extended the VERP, acute amiodarone did not. In the mapped region, acute and chronic amiodarone decreased the number of VF wavefronts by 42% and 60%. Acute amiodarone decreased conduction block by 22%, while chronic amiodarone increased block by 41% but decreased wave fractionation by 50%. Both chronic and acute amiodarone increased the size of the core of re-entrant circuits and decreased the incidence of re-entry by 44% and 57%; however, chronic amiodarone increased wavelength, while acute amiodarone did not.
Conclusions:
Neither acute nor chronic amiodarone change the DFT. While both acute and chronic amiodarone decrease the number of wavefronts, decrease the incidence of re-entry and increase the size of re-entrant cores in the mapped region during VF, they achieve these antiarrhythmic effects through different electrophysiologic mechanisms. Chronic amiodarone prolonged the VF cycle length and slowed conduction velocity, indicating it increased the wavelength and/or the excitable gap.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias