Evaluation of heated humidifiers for use on intubated patients: a comparative study of humidifying efficiency, flow
Jörg Rathgeber1, Stephan Kazmaier, Olaf Penack
1Department of Anaesthesiology and Intensive Care Medicine, Albertinen Hospital, Süntelstrasse 11a, 22457 Hamburg, Germany. joerg.rathgeber@albertinen.de
Objective:
Evaluation of humidification efficiency, flow resistance, and alarm functions of heated humidifiers (HH;(Kendall-Aerodyne-delta, Fisher&Paykel-MR 730; Dräger-Aquapor; Puritan-Bennett-Cascade II) in accordance with ISO/EN-8185:1997 and on a ventilated lung model in accordance with ISO/EN-9360:2000.
Methods:
Humidification efficiency was evaluated by (a) measuring the water content of the inspiratory air on perfusion with different gas flows, (b) measuring the water loss of a lung model, and (c) simultaneous measurement of the in- and expiratory water content with a capacitive hybrid sensor. The resistance characteristics were measured, the data were compared with a mathematical approximation. The alarm functions were determined.
Results:
The humidification efficiency of HHs is a function of gas flow and design characteristics. In HHs with tube heating it is possible to make settings at which the inspiratory humidity falls below the minimal value of 33 mgH(2)O/l stipulated by ISO/EN-8185:1997. The inspiratory resistances extend from 0.5 to 4.4 cmH(2)O l(-1) s(-1); the expiratory flow resistances of the devices are low. The alarm functions of HHs with tube heating are inadequate for cases involving both "dry start" and "running dry."
Conclusions:
Efficiency data that allow a direct comparison with heat and moisture exchangers data according to ISO/EN-9360:2000 can also be determined for HH. HH do not prevent pulmonary water losses in intubated patients. These losses can exceed the physiological range. The airway resistance of the Cascade II prohibits its use in spontaneously breathing patients. The warning and shut-off features of HH are unacceptable and hazardous.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)