Related Experiment Videos
First magnetomyographic recordings of uterine activity with spatial-temporal information with a 151-channel sensor
Hari Eswaran1, Hubert Preissl, James D Wilson
1Graduate Institute of Technology, University of Arkansas for Medical Sciences, Little Rock, 72205, USA.
Objective:
The purpose of this study was to examine the feasibility of recording the spatial-temporal magnetomyographic activity from the pregnant uterus with the use of the newly developed 151-channel noninvasive device, known as the superconducting quantum interference device array for reproductive assessment.
Study Design:
Uterine magnetomyographic signals were recorded from 10 pregnant subjects with the 151-channel sensor array curved to fit the pregnant abdomen. The recording sessions were 16 minutes in length, with a sampling rate of 250 Hz.
Results:
Uterine activity bursts were successfully recorded with the superconducting quantum interference device array for reproductive assessment system. By obtaining a contour plot of the magnetic field distribution, we were able to localize the areas of activation over the uterus during a contraction. Also, it was possible to calculate the time delay in the propagation of the activity across the uterus.
Conclusion:
Using superconducting quantum interference device array for reproductive assessment system, we have established the feasibility of recording uterine contractile activity with spatial-temporal resolution that is high enough to determine the regions of localized activation and propagation over the uterus.