Related Experiment Video
Updated: Aug 12, 2026

Physical Isolation of Endospores from Environmental Samples by Targeted Lysis of Vegetative Cells
Published on: January 21, 2016
Immobilized bacterial spores for use as bioindicators in the validation of thermal sterilization processes
D Serp1, U von Stockar, I W Marison
1Department of Chemistry, Swiss Federal Institute of Technology (EPFL), Lausanne.
Abstract:
Spores of Bacillus subtilis ATCC 6051 and Bacillus stearothermophilus NCTC 10003 were immobilized in monodisperse alginate beads (diameter, 550 microm +/- 5%), and the capacity of the immobilized bioindicators to provide accurate and reliable F-values for sterilization processes was studied. The resistance of the beads to abrasion and heat was strong enough to ensure total retention of the bioindicators in the beads in a sterilization cycle. D- and z-values for free spores were identical to those for immobilized spores, which shows that immobilization does not modify the thermal resistance of the bioindicators. A D(100 degrees C) value of 1.5 min was found for free and immobilized B. subtilis spores heated in demineralized water, skimmed milk, and milk containing 4% fat, suggesting that a lipid concentration as low as 4% does not alter the thermal resistance of B. subtilis spores. Providing that the pH range is kept between 3.4 to 10 and that sufficiently low concentrations of Ca2+ competitors or complexants are present in the medium, immobilized bioindicators may serve as an efficient, accurate, and reliable tool with which to validate the efficiency of any sterilization process. The environmental factors (pH, media composition) affecting the thermoresistance of native contaminants are intrinsically reflected in the F-value, allowing for a sharper adjustment of the sterilization process. Immobilized spores of B. stearothermophilus were successfully used to validate a resonance and interference microwave system that is believed to offer a convenient alternative for the sterilization of temperature-sensitive products and medical wastes.
More Related Videos
10:03Investigating the Detrimental Effects of Low Pressure Plasma Sterilization on the Survival of Bacillus subtilis Spores Using Live Cell Microscopy
Published on: November 30, 2017
13:47Opsono-Adherence Assay to Evaluate Functional Antibodies in Vaccine Development Against Bacillus anthracis and Other Encapsulated Pathogens
Published on: May 19, 2020
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Endospores and Sporulation
Physical Methods for Controlling Microbial Growth: Temperature
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Key Techniques in Microbiology