A Rat Model of Ventricular Fibrillation and Resuscitation by Conventional Closed-chest Technique
Published on: April 26, 2015
Adverse outcomes of interrupted precordial compression during automated defibrillation
Ting Yu1, Max Harry Weil, Wanchun Tang
1Institute of Critical Care Medicine, Palm Springs, Calif 92262-5309, USA.
Background:
Current versions of automated external defibrillators (AEDs) require frequent stopping of chest compression for rhythm analyses and capacity charging. The present study was undertaken to evaluate the effects of these interruptions during the operation of AEDs.
Methods And Results:
Ventricular fibrillation was electrically induced in 20 male domestic swine weighing between 37.5 and 43 kg that were untreated for 7 minutes before CPR was started. Defibrillation was attempted with up to 3 sequential 150-J biphasic shocks, but each was preceded by 3-, 10-, 15-, or 20-second interruptions of chest compression. The interruptions corresponded to those that were mandated by commercially marketed AEDs for rhythm analyses and capacitor charge. The sequence of up to 3 electrical shocks and delays were repeated at 1-minute intervals until the animals were successfully resuscitated or for a total of 15 minutes. Spontaneous circulation was restored in each of 5 animals in which precordial compression was delayed for 3 seconds before the delivery of the first and subsequent shocks but in none of the animals in which the delay was >15 seconds before the delivery of the first and subsequent shocks. Longer intervals of CPR interventions were required, and there was correspondingly greater failure of resuscitation in close relationship to increasing delays. The durations of interruptions were inversely related to the durations of subthreshold levels of coronary perfusion pressure. Postresuscitation arterial pressure and left ventricular ejection fraction were more severely impaired with increasing delays.
Conclusions:
Interruptions of precordial compression for rhythm analyses that exceed 15 seconds before each shock compromise the outcome of CPR and increase the severity of postresuscitation myocardial dysfunction.
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Introduction Cardiac Emergencies
Cardiopulmonary Resuscitation I: Adult
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation IV: Pharmacological Management