Related Experiment Video
Updated: Sep 30, 2026

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Inspiratory pressure-volume curves obtained using automated low constant flow inflation and automated occlusion
Quentin Blanc1, Jean-Michel Sab, François Philit
1Service de Réanimation Médicale et Assistance Respiratoire, Hôpital de la Croix Rousse, 103 Grande Rue de la Croix Rousse, 69004 Lyon, France.
Objective:
To compare the inspiratory volume pressure (VP) curves of the respiratory system (rs) produced by static occlusion (OCC) and dynamic low constant flow inflation (LCFI) methods using a new device in acute respiratory distress syndrome (ARDS) patients.
Setting:
A multidisciplinary 24-bed ICU in a tertiary university hospital.
Patients:
Eleven intubated and mechanically ventilated patients with ARDS.
Measurements And Results:
OCC and LCFI methods were performed using the same ventilator, which had been specifically implemented for this purpose. LCFI of 5, 10, and 15 l/min and OCC were applied in a random order at zero end-expiratory positive pressure. Airway pressure was measured both proximal (P(ao)) and distal (P(tr)) to the endotracheal tube. Lower inflection point (LIP) and maximal slope (C(max,rs)) were estimated using unbiased iterative linear regressions. LIP(rs) was obtained in all patients under LCFI and in nine patients under OCC. With LCFI of 5, 10, 15 l/min and OCC the average LIP(rs) values were 12.2 +/- 3.9, 12.9 +/- 4, 14.3 +/- 3.4, and 11.9 cm H(2)O for P(ao) and 11.9 +/- 3.9, 11.5 +/- 3.3, 12.5 +/- 3.4 and 11.8 +/- 4.4 for P(tr), respectively. Only the mean values of LIP(rs) for P(ao) with LCFI at 15 l/min were significantly different from those obtained for OCC. The C(max,rs) values found with the two methods were similar.
Conclusions:
An LCFI less than or equal to 10 l/min seems to be a quick, safe, and reliable method to determine LIP(rs) and C(max,rs) at the bedside.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...