Related Experiment Video
Updated: Sep 30, 2026

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Endocardial mapping and ablation of tachycardia guided by noncontact balloon catheter mapping system
Jiangang Zou1, Kejiang Cao, Minglong Chen
1Department of Cardiology, The First Affiliated Hospital, Nanjing Medical University, China. jgzou@jlonline.com
Objective:
To describe a new noncontact balloon catheter mapping system and to assess the clinical utility of this system for guiding endocardial mapping and ablation of tachycardia.
Methods:
Five patients with tachycardia underwent endocardial mapping and radiofrequency ablation using the noncontact balloon catheter mapping system. A 9 French, 64-electrode balloon catheter and a conventional 7 French electrode catheter for mapping and ablation were positioned in the same ventricular chamber. Ventricular three-dimensional geometry was established by the computerized mapping system. Using a boundary element inverse solution, 3360 virtual endocardial electrograms were computerized and used to derive isopotential maps. The earliest endocardial activation site, the exit site and the activation sequence of tachycardia or the critical isthmus of the reentry circuit were identified. Radiofrequency ablation with circular or linear lesion was performed at the target sites guided by the locator system.
Results:
Six clinical types of tachycardia, 5 of which were ventricular tachycardia and one was concealed fasciculoventricular fiber mediated tachycardia, were induced by programmed stimulation. The mean cycle length of these tachycardias was 336.6 +/- 42.69 msec. The earliest activation site and the exit site of 5 mapped tachycardias were all identified using the system. One type of ventricular tachycardia was hemodynamically unstable and difficult to terminate, and could not be mapped. Among the 6 types of tachycardias, radiofrequency ablation was successful in 4. There was no complication during and after the procedure. During the mean follow-up of 6 months, no tachycardia recurred in the patients with a successful ablation.
Conclusions:
The noncontact mapping system described in this study has advantage over conventional mapping techniques for refractory tachycardia. It is not only helpful for understanding the electrophysiologic mechanism of a complex case, but also suitable for mapping hemodynamically intolerated and nonsustained ventricular tachycardia.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization III: Left Heart Catheterization