Assessment of three methods of pH probe positioning in preterm infants

Anthony J B Emmerson1, Terri Chant, Judith May

  • 1Neonatal Medical Unit, St. Mary's Hospital, Whitworth Park, Manchester, UK. anthony.mmerson@man.ac.uk

Insights

Accurate pH probe placement in preterm infants is vital for gastroesophageal reflux assessment. Strobel's formula offers a reliable, non-radiographic method for precise pH catheter positioning in this vulnerable population.

Area of Science:

  • Neonatal Medicine
  • Pediatric Gastroenterology
  • Medical Device Technology

Background:

  • Accurate gastroesophageal reflux (GER) assessment in neonates relies on precise pH catheter placement.
  • Very low birth weight infants present unique challenges for pH monitoring due to potential for significant measurement errors from minor positioning inaccuracies.
  • Established methods for verifying catheter placement in infants are often problematic, necessitating investigation into alternative techniques.

Purpose of the Study:

  • To evaluate the accuracy of Strobel's formula, manometry, and the acid-alkali interface for pH probe positioning in preterm infants.
  • To identify a reliable, non-invasive method for confirming pH catheter placement, potentially reducing the need for radiographic confirmation.

Main Methods:

  • Twenty-six preterm infants (<35 weeks gestation) with suspected GER underwent pH monitoring.
  • Probe position was initially assessed using Strobel's formula, manometry, and the acid-alkali interface.
  • Final confirmation of probe placement was achieved through chest x-ray.

Main Results:

  • A strong positive correlation was observed between pH probe position calculated by Strobel's formula and x-ray confirmation (p<0.001).
  • A significant positive correlation was found between the acid-alkali interface method and x-ray positioning.
  • Manometry showed no correlation with x-ray position and proved difficult to apply in this infant cohort.

Conclusions:

  • Strobel's formula is a suitable and accurate method for positioning pH probes in preterm infants, potentially eliminating the need for routine chest x-rays.
  • The acid-alkali interface serves as a reliable secondary method for confirming pH probe position when calculable, complementing Strobel's formula.