Related Experiment Video
Updated: Aug 4, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Blood glucose overestimation in diabetic patients on continuous ambulatory peritoneal dialysis for end-stage renal
Samson O Oyibo1, G M Pritchard, L McLay
1Department of Medicine, Manchester Royal Infirmary, Oxford Road, Manchester M13 9WL, UK. samson@dc.cmht.nwest.nhs.uk
Aims:
Diabetic patients on continuous ambulatory peritoneal dialysis (CAPD) for renal failure depend on glucose analysers for regular monitoring of glycaemic control. We aim to inform health professionals of the potentially dangerous overestimation of blood glucose values by some analysers in patients using Icodextrin for dialysis.
Methods:
Twenty-five patients on continuous ambulatory peritoneal dialysis (10 patients on an 8-12-h nocturnal exchange of Icodextrin) had random glucose analysis performed on venous blood using standardized reference laboratory (lab) technique (glucose oxidase GOD-PAP), and simultaneously on capillary blood using the Precision Q.I.D System (glucose oxidase method) and the Advantage meter (glucose dehydrogenase method).
Results:
The Precision Q.I.D System agreed with the lab results in both the Icodextrin group and the non-Icodextrin group (80-90% of values fell within 20% of the corresponding lab result). In contrast, the Advantage meter agreed with the lab results only in the non-Icodextrin group (95% of values within 20% of the corresponding lab value), and not in the Icodextrin group, where only 5% of the analyser values fell within 20% of the corresponding lab value.
Conclusions:
The Precision Q.I.D System, which utilizes glucose oxidase reaction, is safe for use in diabetic patients treated with Icodextrin. All analysers must be cross-checked with the laboratory reference method before use in these patients.
Related Concept Videos
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Hypoglycemia and Glucagon
Peritoneal Dialysis I: Introduction and Procedure
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Hyperglycemia
Diabetic Nephropathy