Trevor W Dawson1, Krzysztof Caputa, Maria A Stuchly
1Department of Electrical and Computer Engineering, University of Victoria, BC, Canada. tdawson@ece.uvic.ca
Contact currents can interfere with cardiac pacemakers. This study found that 60-Hz contact currents can disrupt unipolar pacemakers at 24–45 microA and bipolar pacemakers at 63–340 microA.
You might also read
Articles linked to this work by shared authors, journal, and citation graph.
Area of Science:
Background:
Purpose of the Study:
Main Methods:
Main Results:
Conclusions: