Related Experiment Video
Updated: Sep 30, 2026

Simple In-House Ultra-High Performance Capillary Column Manufacturing with the FlashPack Approach
Published on: December 4, 2021
Repeatability and reproducibility of retention data and band profiles on six batches of monolithic columns
Marianna Kel1, Georges Guiochon
1Department of Chemistry, The University of Tennessee, Knoxville 37996-1600, USA.
Abstract:
Chromatographic data were acquired for eight different mixtures, under five different sets of experimental conditions, for a total of 30 neutral, acidic and basic test compounds, on a series of six Chromolith Performance columns from Merck. These columns are made of a C18 chemically bonded silica monolith. Each column belonged to a different production batch, so the data reported here characterize their batch-to-batch reproducibility. The parameters studied in this work were the retention times, the retention and separation factors, the hydrophobic and the steric selectivities, the column efficiencies, and the tailing factors for all 30 compounds.
Related Concept Videos
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Chromatographic Methods: Terminology
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Principles Of Column Chromatography
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...