Related Experiment Videos
Mass transfer in blood oxygenators using blood analogue fluids
S Ranil Wickramasinghe1, Carin M Kahr, Binbing Han
1Department of Chemical and Bioresource Engineering, Colorado State University, Fort Collins, Colorado 80523-1370, USA. wickram@engr.colostate.edu
Abstract:
Mass transfer correlations for hollow fiber blood oxygenators have been determined experimentally using Newtonian and non-Newtonian blood analogue fluids. The Newtonian fluids consisted of deionized water and glycerol/water mixtures. The non-Newtonian fluids were prepared by adding small amounts of xanthan gum to the Newtonian blood analogue fluids. The rheological behavior of the non-Newtonian blood analogue fluids was modeled using the power law. The diffusion of oxygen into and out of the Newtonian and non-Newtonian blood analogue fluids has been studied. The liquid stream flowed outside and across bundles of woven hollow fibers, while the gas stream flowed inside the fibers.