Assessment of cardiac output from systemic arterial pressure in humans
Salvatore M Romano1, Massimo Pistolesi
1Department of Critical Care, University of Florence, Italy. m.romano@dac.unifi.it
Objective:
To evaluate the reliability, by comparison with established techniques, of a new method to assess cardiac output, called pressure recording analytical method (PRAM), deriving from the analysis of the arterial pressure profile in the time domain the arterial-pressure-blood flow relationship.
Design:
Criterion standard.
Setting:
Hemodynamics laboratory at an university medical center.
Patients:
Twenty-two hemodynamically stable cardiac patients scheduled for diagnostic right and left heart catheterization.
Interventions:
None.
Measurements And Main Results:
Cardiac index was simultaneously estimated by direct-oxygen Fick method, thermodilution, and PRAM applied to pressure signals recorded either invasively from an aortic catheter (PRAMa) or noninvasively at the finger (PRAMf) by photoplethysmography. Cardiac index values obtained by established techniques were significantly correlated with those estimated by PRAM: Fick method vs. PRAMa, r(2) =.88, vs. PRAMf, r(2) =.94; thermodilution vs. PRAMa, r(2) =.77, vs. PRAMf, r(2) =.77. The Bland-Altman analysis showed agreement between the Fick method and PRAM, with all data points comprised within the limits of agreement (+/-2SD) (mean difference +/- SD: -0.012 +/- 0.187 L x min(-1) x m(-2) for PRAMa; 0.024 +/- 0.167 L x min(-1) x m(-2) for PRAMf). Agreement was also found between thermodilution and PRAM, with all but one data point lying within the limits of agreement (mean difference +/- SD: -0.154 +/- 0.348 L x min(-1) x m(-2) for PRAMa; -0.108 +/- 0.348 L x min(-1) x m(-2) for PRAMf).
Conclusions:
In the range evaluated (cardiac index from 1.65 to 3.91 L x min(-1) x m(-2) by the Fick method), PRAM provides reliable invasive and noninvasive estimates of cardiac output in hemodynamically stable cardiac patients. PRAM may prove clinically useful for the beat-to-beat monitoring of cardiac output.
Related Concept Videos
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Pulse
The pulse serves as a clinical indicator...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...