Continuous blood gas monitoring using an in-dwelling optode method: comparison to intermittent arterial blood gas
K Rais-Bahrami1, Oswaldo Rivera, Gerald T Mikesell
1Department of Neonatology, Children's National Medical Center and The George Washington University School of Medicine, Washington, DC 20010, USA.
Introduction:
The ability to measure postmembrane arterial blood gases is essential in the management of critically ill neonates treated with extracorporeal membrane oxygenation (ECMO). A new technology using, the Paratrend 7 system (Diametrics Medical, High Wycombe,UK) allows for continuous measurement of pH, PCO(2) and PO(2), and calculates oxygen saturation, bicarbonate, and base excess.
Objective:
To evaluate and compare the results of continuous blood gas measurement using the Paratrend 7 system with a standard system of blood gas analysis in our intensive care unit.
Design:
Prospective, controlled, interventional study.
Setting:
The neonatal intensive care unit of a tertiary referral center.
Patients:
Neonates who required extracorporeal life support and were expected to have frequent postmembrane arterial blood sampling during the testing period.
Research Design And Method:
To enable Paratrend 7 sensor access to the ECMO circuit, the postmembrane access port extension set that is routinely used for blood drawn for blood gas analysis was used. The study began with the insertion of the Paratrend 7 sensor. Subjects remained on the study until the ECMO was discontinued and/or frequent blood gases were no longer needed. The blood gas results from the Paratrend 7 system were not used in clinical management of the patient. BLOOD GAS MEASUREMENT: During the study period, with each blood sample drawn for laboratory analysis, a printout from the Paratrend 7 monitor was recorded for comparison.
Results:
A total of 242 pairs of blood gas samples were collected from 10 neonates. The mean bias/precision for pH was -0.02/0.04; for PO(2) 68.35/93.44 mm Hg; and for PCO(2) 1.75/4.23 mm Hg. The correlation (r value) between the sensor reading and the blood gases were 0.89 for pH, 0.96 for PO(2), and 0.73 for PCO(2) (Table 1).
Conclusion:
The blood gases compared in the two methods had a strong correlation for pH, PCO(2) and PO(2). Results of this study indicate that this technology provides an accurate means of monitoring continuous blood gas parameters in neonatal ECMO patients. Use of the Paratrend 7 should allow reduced health-care provider exposure to blood and decreased patient iatrogenic blood loss.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...