Purpose:
To assess the ability of various facets of embryo culture (microscope stage warmers, volumes of culture media, culture vessel lids, and type of culture incubator) to maintain a constant temperature in vitro.
Methods:
Ability to maintain 37.0 degrees C in the microenvironment of gametes was recorded by digital thermocouple in the chosen facets of in vitro culture.
Results:
Stage warmers are highly variable in their ability to maintain the set temperature (range 33.8 degrees C-37.0 degrees C after 60 s). Temperature loss in culture media is both volume and vessel dependent, and the direct heat transfer culture incubator (MINC) has superior temperature maintenance compared with a large volume air convection incubator (FORMA), where temperature regain from 35.0 degrees C to 37.0 degrees C took 5.5 min compared to >20 min.
Conclusions:
There are large measurable differences in the ability to maintain set temperature that depend on the stage warmer used, volume of media, use of vessel lids, and the type of incubator chosen for IVF culture.
Related Concept Videos
Key Techniques in Microbiology
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Physical Methods for Controlling Microbial Growth: Temperature
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Factors Influencing Microbial Growth: Temperature
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...