False positive blood tacrolimus concentration in microparticle enzyme immunoassay
Masato Homma1, Takashi Tomita, Kenji Yuzawa
1Department of Pharmaceutical Sciences, University of Tsukuba, Ibaraki, Japan. masatoh@mail.hosp.tsukuba.ac.jp
Abstract:
The difference in the blood concentration of tacrolimus between the assay methods, microparticle enzyme immunoassay (MEIA) and enzyme linked immunosorbent assay (ELISA) was observed in a liver transplant recipient with anemia. MEIA provided significantly higher concentration than those of ELISA (7.8+/-1.9 vs. 5.0 +/- 1.8nglml, p<0.05) while the patient had low haematocrit <25%. The difference, however, was not observed during the periods with haematocrit >25%. This observation suggested that unknown tacrolimus levels generated from difference in assay methods gave incorrect blood tacrolimus during anemia. False positive concentration of tacrolimus ranging 0.1-3.3 ng/ml was observed in MEIA applying to the blood samples obtained from the patients without receiving tacrolimus. The false positive tacrolimus increased in the samples with lower hematocrit, suggesting that MEIA gave incorrect blood tacrolimus during anemia. Since MEIA potentially overestimates the tacrolimus levels, ELISA should be used for blood tacrolimus monitoring in the patients with anemia.
More Related Videos
06:08Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
07:37Microwaving and Fluorophore-Tyramide for Multiplex Immunostaining on Mouse Adrenals − Using Unconjugated Primary Antibodies from the Same Host Species
Published on: February 21, 2020