Background:
The use of sphygmomanometers may lead to problems in investigations on health of young people. The purpose of this paper is to present the validation of the blood pressure (BP) collected during the survey 'Youth Heart Health' in Hainaut by using second sample of young people in Hainaut for which BP was measured by a manual taking of tension and by an electronic device.
Methods:
Validation was done with a control sample of 343 young with five successive BP measures: twice with the mercury sphygmomanometer and three with DXL. We compared the manual and the electronic measures in order to study the correlation between the two methods. The control sample was used in order to compare the BP measurements with the results of the survey on the health of young people in Hainaut.
Results:
The differences between manual systolic BP and Dinamap measures are significant (differences in averages 3.6 mmHg; d.s. 7.8; 95% CI: 2.8-4.4 mmHg; p < 0.001) and the regression coefficient is -0.015. Diastolic BP is significantly higher with the manual method than with Dinamap (differences in means: 8.2 mmHg; d.s. 7.0; 95% CI: 7.4-8.9 mmHg; p < 0.001) and the coefficient of regression is 0.096. We observe a difference in BP between the general survey and the control group (10.3 mmHg for systolic manual control and of 10.9 mmHg for systolic electronic control; 3.3 mmHg for diastolic manual control and of 11.5 mmHg for diastolic electronic control).
Conclusions:
The values of BP of the 'Youth Heart Health' are significantly higher. These observations indicate the difficulties in the choice of the tool for measurements of BP in epidemiological investigations in the teenagers. In order to decrease skews of observations in the measurement of BP, it is recommended to use a valid electronic instrument. Dinamap XL is an instrument of choice in such studies of young people. Environmental and organisational factors may also explain the observed difference, which means that survey's protocols should address this issue carefully.
Related Concept Videos
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: