Related Experiment Video
Updated: Sep 29, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Relationship of neonatal endotracheal tube size and airway resistance
Melisa J Oca1, Michael A Becker, Ronald E Dechert
1Department of Pediatrics, Division of Neonatal-Perinatal Medicine, Unviersity of Michigan Health System, Ann Arbor, Michigan, USA.
Background:
Infants receiving mechanical ventilation require narrow-lumen, small-diameter endotracheal tubes.
Objective:
Compare the resistances of endotracheal tubes used in the neonatal intensive care unit.
Methods:
Endotracheal tubes of internal diameter 2.5, 3.0, 3.5, and 4.0 mm were tested with a standard neonatal ventilator and a test lung. An endotracheal tube of each diameter was cut to 12 cm and connected to a flow transducer at one end and the test lung at the other. Serial measurements of resistance were made at various flows (6, 8, 10, and 12 L/min) and ventilator rates (30-90 breaths/min) encompassing the ranges of clinical practice. Analysis of variance was performed for each tube size, comparing resistance to flows and ventilator rates.
Results:
Resistance was significantly higher with the 2.5 mm tube than with the others. There was also a consistent trend, in all the tube sizes, towards higher resistance as flow was increased.
Conclusions:
The higher resistance of the 2.5 mm tube may be detrimental to extremely low birthweight infants kept on mechanical support merely "to grow." The higher resistance may increase the work of breathing and thus increase caloric expenditure and impede growth.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...