An alignment method for mammographic X-ray spectroscopy under clinical conditions
S Miyajima1, K Imagawa, M Matsumoto
1Graduate School of Medicine, Course of Health Sciences, Osaka University, 1-7 Yamadaoka, Suita, Osaka, 565-0871, Japan.
The British Journal of Radiology
|August 30, 2002
Abstract:
This paper describes an alignment method for mammographic X-ray spectroscopy under clinical conditions. A pinhole, a fluorescent screen, a laser device and the case for a detector are used for alignment of the focal spot, a collimator and a detector. The method determines the line between the focal spot and the point of interest in an X-ray field radiographically. The method allows alignment for both central axis and off-axis directions.