Evaluation of an oscillometric blood pressure monitor for use in anesthetized cats
Objective:
To determine accuracy of an oscillometric blood pressure monitor used over a wide range of pressures in anesthetized cats.
Design:
Prospective study.
Animals:
6 healthy cats.
Procedure:
4 female cats and 2 male cats that weighed 2.7 to 4.5 kg (5.9 to 9.9 lb) and were 2 to 8 years old were anesthetized. Blood pressure was measured directly with an arterial catheter placed in the right femoral artery and indirectly from the left antebrachium by use of an oscillometric monitor. A series of diastolic arterial pressure (DAP), mean arterial pressure (MAP), and systolic arterial pressure (SAP) measurements were obtained during hypotension, normotension, and hypertension. Values obtained indirectly and directly were compared.
Results:
The oscillometric monitor was accurate for DAP and MAP throughout the entire pressure range and met the standards of the Association for the Advancement of Medical Instrumentation (mean +/- SD difference from values obtained directly, < or = 5 +/- 8 mm Hg). The SAP was increasingly underestimated with increasing overall pressure; mean differences from direct measurements were -5.2, -12.1, and -17.7 mm Hg during hypo-, normo-, and hypertension, respectively. Standard deviations for SAP were all < or = 8 mm Hg. The monitor gave readings during all attempts. The direct blood pressure recording system appeared to perform well with neither under- nor overdamping.
Conclusions And Clinical Relevance:
Except for a minor underestimation of SAP during normo- and hypertension, the oscillometric monitor yielded reliable and easily obtainable blood pressure measurements in anesthetized cats.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)