Optimization of the composition of phantom materials for computed tomography
Related Experiment Video
Updated: Aug 12, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
P Homolka1, A Gahleitner, M Prokop
1Department of Biomedical Engineering and Physics, University of Vienna, Austria. peter.homolka@univie.ac.at
Abstract:
Using available data for photon attenuation and tissue composition, a computer code was developed for the optimization of the composition of phantom materials for diagnostic radiology. The code allows selection of attenuation data in a photon energy range from 1 to 150 keV and the choice of a suitable weight function in the energy interval chosen. For applications in CT imaging a weight function is available reflecting the contribution of the x-ray spectrum to the CT-signal. Several phantom materials for CT were optimized (body fat, trabecular bone, an average bone composition for C4 vertebrae and water) by varying the mineral components in a polymer base in order to adjust x-ray attenuation properties. Measurements with the water equivalent material (PSPP1) showed good agreement of calculated and measured HU values (AHU = 7.3 +/- 5.3 at 80 kVp and 4.0 +/- 2.7 at 140 kVp) and little variation of HU for tube voltages from 80 to 140 kVp. The method provides a fast and flexible means for obtaining optimized phantom materials for a large variety of tissue compositions and energy ranges.
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography