Related Experiment Video
Updated: Sep 29, 2026

Generation of Electronic Cigarette Aerosol by a Third-Generation Machine-Vaping Device: Application to Toxicological Studies
Published on: August 25, 2018
The comparative risks of bacterial contamination between a venturi atomizer and a positive displacement atomizer
Timothy R Wolfe1, Todd A Hillman, Philip J Bossart
1Department of Surgery, University of Utah School of Medicine, USA.
Introduction:
This laboratory study determined the incidence of internal contamination of Venturi principle atomizers and positive displacement atomizers exposed to high external concentrations of Staphylococcal aureus (Staph).
Methods:
Atomizer device nozzle tips were immersed into a Staph solution and I ml of spray was atomized via compressed wall air (Venturi) or hydraulic pump (positive displacement). The Venturi nozzle was then wiped with 70% isopropyl alcohol while the disposable positive displacement nozzle was replaced. After 30 minutes, 1 ml of atomized fluid was collected and cultured and the process was repeated. After sixteen uses the fluid remaining in the bottles was cultured. The Venturi atomizer also was subjected to a single use trial to determine the location of device contamination.
Results:
Venturi atomizers sprays grew Staph in every case (144/144), while positive displacement atomizer sprays never grew contaminants (0/144; p < 0.0001). At the end of 16 uses, 7/9 of Venturi atomizers had Staph within their medication reservoirs while none (0/9; p = 0.002) existed in the positive displacement atomizers. After a single use of the Venturi atomizer, the medication reservoir, the air lumen and the medication lumen of the nozzle were all contaminated with Staph.
Conclusions:
External bacterial contamination of the atomizer nozzle tip results in internal bacterial contamination of Venturi devices in as little as one use but not of positive displacement devices. These results warrant further investigation to determine whether a risk of cross-contamination exists in a clinical setting.
Related Concept Videos
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Stimulants
Cocaine can be administered via snorting, injection, or smoking. It primarily functions by blocking the reuptake of dopamine, resulting in a euphoric high characterized by an intense sensation of happiness and...
Atomic Absorption Spectroscopy: Atomization Methods
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...